Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SmartPath to dStream for 1.5T - For use as a Recalled by Philips North America Llc Due to Integrated Radio Frequency (IRF) Carrier assembly screws may...

Date: September 3, 2024
Company: Philips North America Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

SmartPath to dStream for 1.5T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782112; 2) 781260

Quantity: 32 units

Why Was This Recalled?

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected