Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

SENSE XL TORSO COIL 1.5T. 16-element receive only coil for Recalled by Philips North America Llc Due to Potential for coils to heat up and harm...

Date: June 5, 2024
Company: Philips North America Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

SENSE XL TORSO COIL 1.5T. 16-element receive only coil for torso and abdomen imaging in Philips MR Intera and Achieva Systems. Model Numbers: 453567141882, 453567141883. Coils may be included in kits with Mode Numbers: 989603014351 & 989603014352.

Quantity: 729 units

Why Was This Recalled?

Potential for coils to heat up and harm patients (burn).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips North America Llc

Philips North America Llc has 301 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report