Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Recalled by Philips North America Due to Issue 1: Mixed Images: The Radio Fluoroscopy (RF)...

Date: December 15, 2023
Company: Philips North America
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America directly.

Affected Products

ProxiDiagnost N90- A Multi-functional general Radiography and Fluoroscopy (R/F) system Ref: (1)706100 (2)706110

Quantity: 151 units; 138 units US; 13 units OUS

Why Was This Recalled?

Issue 1: Mixed Images: The Radio Fluoroscopy (RF) viewer may also display a previous patient s radiography (RAD) images when starting the next patient scan while the previous patient image export is still processing. If the issue occurs, there is a potential for incorrect patient data to be displayed in the image. Issue 2: Detector Access point: Security vulnerability specific to the Wireless Portable Detector configuration items in Philips Support Connect (PSC). Due to this vulnerability it is possible, with physical access to the system and knowledge of specific settings, to modify and export data to removable media (example: USB)

Where Was This Sold?

This product was distributed nationwide across all 50 states.