Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Philips SureSigns VM4 Recalled by Philips North America, LLC Due to Fail to Comply with Chinese Standard YY1079-2008: Clauses...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.
Affected Products
Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only) SureSigns VM4 (863063, 863085) SureSigns VM6 (863064, 863065, 863086) SureSi"gns VM8 (863066, 863068, 863087, 863088) Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421 453564024441,453564024461, 453564024481
Quantity: 31773 units
Why Was This Recalled?
Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit
Where Was This Sold?
International distribution in the country of China.
About Philips North America, LLC
Philips North America, LLC has 88 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report