Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips DigitalDiagnost C50 -intended for use in generating radiographic images Recalled by Philips North America Llc Due to If the PM for the telescopic carriage is...

Date: March 9, 2023
Company: Philips North America Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201

Quantity: US: 10 units; OUS: 570 units

Why Was This Recalled?

If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users

Where Was This Sold?

This product was distributed nationwide across all 50 states.