Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Azurion Interventional Fluoroscopic X-Ray System Recalled by Philips North America Llc Due to When the user presses or releases both the...

Date: September 24, 2021
Company: Philips North America Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)

Quantity: 428 (US); 581 (OUS)

Why Was This Recalled?

When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.

Where Was This Sold?

This product was distributed nationwide across all 50 states.