Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IntelliVue MX700 patient monitor Recalled by Philips North America, LLC Due to Some Intellivue MX700/MX800 Patient Monitors were delivered with...

Date: June 30, 2020
Company: Philips North America, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.

Affected Products

IntelliVue MX700 patient monitor, model no. 865241

Quantity: 28

Why Was This Recalled?

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.

Where Was This Sold?

This product was distributed to 5 states: CA, HI, MA, NY, UT

Affected (5 states)Not affected

About Philips North America, LLC

Philips North America, LLC has 88 total recalls tracked by RecallDetector.