Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ingenia Elition X Magnetic Resonance (MR) system Recalled by Philips North America Due to The system operator is able to bypass the...

Date: June 12, 2023
Company: Philips North America
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America directly.

Affected Products

Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)

Quantity: 640 total

Why Was This Recalled?

The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips North America

Philips North America has 107 total recalls tracked by RecallDetector.