Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use Recalled by Philips North America Llc Due to Pressure relief valve 42 psi installed in the...

Date: July 30, 2021
Company: Philips North America Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS)

Quantity: 11 units

Why Was This Recalled?

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Where Was This Sold?

This product was distributed nationwide across all 50 states.