Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

(1) Ingenia Ambition X (Product Number 781356) Recalled by Philips North America Due to The sealed magnet will experience uncontrolled shutdown known...

Date: July 15, 2021
Company: Philips North America
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America directly.

Affected Products

(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)

Quantity: 56 US; 169 ROW

Why Was This Recalled?

The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)

Where Was This Sold?

This product was distributed to 23 states: CA, DE, FL, GA, IL, KY, ME, MA, MI, MO, NH, NJ, NM, NY, OH, PA, SC, TN, TX, VT, VA, WI, WY

Affected (23 states)Not affected

About Philips North America

Philips North America has 107 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report