Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Incisive CT - Computed Tomography X-Ray System intended to produce Recalled by Philips North America Llc Due to Metal mounting box on the rotating scanner on...

Date: April 15, 2024
Company: Philips North America Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728144

Quantity: 155 units

Why Was This Recalled?

Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

Where Was This Sold?

This product was distributed nationwide across all 50 states.