Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

HeartStart XL+ Defibrillator/Monitor Recalled by Philips North America, LLC Due to The rotary therapy selector switch may fail, resulting...

Date: October 16, 2019
Company: Philips North America, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.

Affected Products

HeartStart XL+ Defibrillator/Monitor, Model 861290

Quantity: 23,749 Units

Why Was This Recalled?

The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from the setting selected by the user.

Where Was This Sold?

This product was distributed nationwide across all 50 states.