Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EasyDiagnost Eleva DRF Recalled by Philips North America, LLC Due to Thermo switches in the main power supply for...

Date: June 2, 2020
Company: Philips North America, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.

Affected Products

EasyDiagnost Eleva DRF, model no. 706050 - Product Usage: X-ray system, diagnostic, general-purpose.

Quantity: 52

Why Was This Recalled?

Thermo switches in the main power supply for the system may be incorrectly installed, resulting in a malfunction.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips North America, LLC

Philips North America, LLC has 88 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report