Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Recalled by Philips North America Llc Due to Significant wear of the Monitor Bracket Assembly shaft...

Date: May 1, 2021
Company: Philips North America Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.

Quantity: 4 units

Why Was This Recalled?

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Where Was This Sold?

This product was distributed nationwide across all 50 states.