Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Allura Exper FD Series. System codes 722003 722010 722022 722005 Recalled by Philips North America, LLC Due to A capacitor inside the converter of the Velara...

Date: December 30, 2019
Company: Philips North America, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.

Affected Products

Allura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001 722002 722006 722012 722008 722013 722020 722025 722015 722023 722038 889006 889016 889009 722005 The device is an X-Ray Generator.

Quantity: 4938

Why Was This Recalled?

A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips North America, LLC

Philips North America, LLC has 88 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report