Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

728327 Ingenuity CT Upgrades-Computed Tomography X-ray system Recalled by Philips North America Llc Due to When setting patient weight unit preferences to pounds,...

Date: June 8, 2021
Company: Philips North America Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America Llc directly.

Affected Products

728327 Ingenuity CT Upgrades-Computed Tomography X-ray system

Quantity: 3 units

Why Was This Recalled?

When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.