Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips MR Systems: Asset 0.5T Recalled by Philips Medical Systems (Cleveland) Inc Due to There is a potential for water to collect...

Date: August 7, 2013
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Philips MR Systems: Asset 0.5T, Apollo 0.5T, Infinion 1.5T, Eclipse/Polaris 1.5T, Panorama 0.6T 1.0T, Magnetic Resonance Imaging Systems intended for clinical/radiologic diagnostic imaging applications.

Quantity: Total 511 (Asset 0.5T,: 35 units Apollo 0.5T: 15 units, Infinion 1.5T: 116 units, Eclipse/Polaris 1.5T: 290 units, Panorama 0.6T: 55 units)

Why Was This Recalled?

There is a potential for water to collect in the vent pipe elbow in the magnet venting system. Water may freeze, blocking the venting system.

Where Was This Sold?

This product was distributed nationwide across all 50 states.