Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A Recalled by Philips Medical Systems, Inc. Due to 1. Device will perform the weekly automated tests...

Date: November 19, 2014
Company: Philips Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems, Inc. directly.

Affected Products

Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A

Quantity: 1553 units

Why Was This Recalled?

1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication