Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). Recalled by Philips Medical Systems North America Inc. Due to Philips is recalling HeartStart Home and OnSite (HS1)...

Date: February 6, 2014
Company: Philips Medical Systems North America Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems North America Inc. directly.

Affected Products

Philips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). Product Usage: The models M5066A, M5067A, and M5068A are designed for use on a person in sudden cardiac arrest who is: -Unresponsive when shaken -Not breathing normally If in doubt, apply the pads. A training class in CPR/AED use is recommended for anyone who may use the HS1. If the person is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child s exact age or weight. If in doubt, use the adult pads. Models M5066A (for HS1 Onsite), M5067A (OnSite) , and M5068A (for HS1 Home).

Quantity: 5075 units total (2724 units in the US and 2351 units outside the US)

Why Was This Recalled?

Philips is recalling HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs) because one or both contact pins may be contaminated by residue from the soldering process. This contamination could prevent an adequate connection between the pads cartridge and the AED. If poor contact between the pads cartridge and AED occurs, the device may be unable to deliver therapy in an emergency.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Medical Systems North America Inc.

Philips Medical Systems North America Inc. has 5 total recalls tracked by RecallDetector.