Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Due to Philips Azurion and Allura systems configured with a...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. directly.
Affected Products
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.
Quantity: 1,861 units
Why Was This Recalled?
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 154 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report