Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Intellispace Portal software Product Usage: Intellispace Portal is a multimodality Recalled by Philips Medical Systems (Cleveland) Inc Due to In the Multi-Modality Tumor Tracking application, when changing...

Date: May 22, 2012
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Intellispace Portal software Product Usage: Intellispace Portal is a multimodality thin-client applications server that delivers full diagnostic viewing and clinical applications to the enterprise. Intellispace Portal is a medical software system that allows multiple users to remotely access the system from compatible computers on a network.

Quantity: 44

Why Was This Recalled?

In the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.

Where Was This Sold?

This product was distributed nationwide across all 50 states.