Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography Recalled by Philips Medical Systems (Cleveland) Inc Due to One of the two wire channels mounted inside...

Date: November 1, 2017
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Ingenuity TF PET/CT - a integrated diagnostic X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) system

Quantity: 2

Why Was This Recalled?

One of the two wire channels mounted inside the Gantry Separation Unit (GSU) that is also used to secure the top two GSU covers, was missing mounting hardware.

Where Was This Sold?

One domestic and 1 foreign consignee ( Spain)