Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ingenia Elition S Recalled by Philips Medical Systems Nederlands Due to During automatic tabletop movement, pressing and holding the...

Date: December 3, 2018
Company: Philips Medical Systems Nederlands
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems Nederlands directly.

Affected Products

Ingenia Elition S, Ingenia Elition X

Quantity: 35

Why Was This Recalled?

During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.

Where Was This Sold?

This product was distributed to 7 states: CA, MN, NJ, NY, OH, TN, TX