Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

iCT SP Model # 728311 Recalled by Philips Medical Systems (Cleveland) Inc Due to In certain instances when performing a cardiac step...

Date: June 24, 2019
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

iCT SP Model # 728311, computed tomography x-ray system

Quantity: 27

Why Was This Recalled?

In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied

Where Was This Sold?

This product was distributed nationwide across all 50 states.