Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brilliance 6 Slice (Air) - Model no. 728256 Recalled by Philips Medical Systems (Cleveland) Inc Due to Closing the Emergency Stop may disable the horizontal...

Date: February 8, 2018
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Brilliance 6 Slice (Air) - Model no. 728256, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.

Quantity: 1

Why Was This Recalled?

Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Medical Systems (Cleveland) Inc

Philips Medical Systems (Cleveland) Inc has 313 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report