Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brilliance iCT Computed Tomography X-ray system Recalled by Philips Medical Systems (Cleveland) Inc Due to A calibrated torque wrench malfunctioned at the supplier...

Date: April 13, 2015
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Brilliance iCT Computed Tomography X-ray system

Quantity: 2

Why Was This Recalled?

A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.

Where Was This Sold?

Internationally to the countries of China and Norway.