Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brilliance iCT Recalled by Philips Medical Systems (Cleveland) Inc Due to Philips has identified that the compensator within the...

Date: July 3, 2018
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories

Quantity: 2

Why Was This Recalled?

Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.

Where Was This Sold?

Distributed only to China and UK.

About Philips Medical Systems (Cleveland) Inc

Philips Medical Systems (Cleveland) Inc has 313 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report