Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brilliance CT 10 Air. Computed Tomography X-ray systems intended to Recalled by Philips Medical Systems (Cleveland) Inc Due to The firm was informed that while raising the...

Date: March 16, 2015
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

Brilliance CT 10 Air. Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Quantity: 153 units total

Why Was This Recalled?

The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.

Where Was This Sold?

Distributed in China.

About Philips Medical Systems (Cleveland) Inc

Philips Medical Systems (Cleveland) Inc has 313 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report