Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BrightView X Recalled by Philips Medical Systems (Cleveland) Inc Due to During collimator exchange, the alignment of the collimator...

Date: December 15, 2017
Company: Philips Medical Systems (Cleveland) Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.

Affected Products

BrightView X, Model #882478 Product Usage: BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only).

Quantity: 1387 in total

Why Was This Recalled?

During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Medical Systems (Cleveland) Inc

Philips Medical Systems (Cleveland) Inc has 313 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report