Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
BrightView X 3/4 model number 882479 BrightView and BrightView X Recalled by Philips Medical Systems (Cleveland) Inc Due to The brake hub was not engaging with the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems (Cleveland) Inc directly.
Affected Products
BrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightView XCT is a gamma camera for Single Photon Emission Computed Tomography (SPECT) and integrates with an attenuation device consisting of flat panel x-ray imaging components. BrightView XCT produces non-attenuation corrected SPECT images and attenuation corrected SPECT images with an x-ray transmission data that may also be used for scatter correction. The nuclear medicine images and BrightView XCT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide anatomical localization of the nuclear medicine data. The BrightView XCT imaging system should only be used by trained healthcare professionals.
Quantity: 0
Why Was This Recalled?
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
Where Was This Sold?
This product was distributed to 34 states: AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NJ, NY, NC, OH, OR, PA, SD, TN, TX, VA, WA, WV, WI
About Philips Medical Systems (Cleveland) Inc
Philips Medical Systems (Cleveland) Inc has 313 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report