Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Allura Xper FD20 Biplane OR Table Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Due to The BIOS battery may deplete faster than initially...

Date: September 3, 2025
Company: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. directly.

Affected Products

Allura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020, (2) 722025; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Quantity: 4 units (2 US, 2 OUS)

Why Was This Recalled?

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Where Was This Sold?

This product was distributed nationwide across all 50 states.