Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Allura Xper FD10 Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Due to Philips has identified that in a limited number...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. directly.
Affected Products
Allura Xper FD10; Model Number: 722010;
Quantity: 56 units
Why Was This Recalled?
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 154 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report