Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system Recalled by PHILIPS HOME HEALTHCARE SOLUTION Due to After the user selects the Lock-in command, the...

Date: December 8, 2020
Company: PHILIPS HOME HEALTHCARE SOLUTION
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PHILIPS HOME HEALTHCARE SOLUTION directly.

Affected Products

ProxiDiagnost N90 Image-intensified fluoroscopic x-ray system

Quantity: 91

Why Was This Recalled?

After the user selects the Lock-in command, the kV and mA values are not locked immediately and can adjust if the collimator shutter position is changed directly after the Lock-in command.