Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Recalled by Philips Healthcare Inc. Due to When the system is switched off or when...

Date: October 19, 2011
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide

Quantity: 11 - US

Why Was This Recalled?

When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Healthcare Inc.

Philips Healthcare Inc. has 36 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report