Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system Recalled by Philips Healthcare Inc. Due to If a lead ruler is not positioned in...

Date: May 11, 2012
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Quantity: 6

Why Was This Recalled?

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

Where Was This Sold?

6 US customers

About Philips Healthcare Inc.

Philips Healthcare Inc. has 36 total recalls tracked by RecallDetector.