Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642 Recalled by Philips Healthcare Inc. Due to Inclination switch in the stand may have been...

Date: July 11, 2012
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips Easy Diagnost Classic-Stationary x-ray system Catalog Numbers: 70642, 70643, 706030, 706031, 706032 Stationary x-ray system, Universal RF system for general use.

Quantity: 12 units

Why Was This Recalled?

Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.

Where Was This Sold?

This product was distributed to 9 states: IN, MA, MI, MO, NY, NC, PA, SC, WI

Affected (9 states)Not affected

About Philips Healthcare Inc.

Philips Healthcare Inc. has 36 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report