Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic Recalled by Philips Healthcare Inc. Due to Creepage Distance between the two primary phases in...

Date: August 14, 2012
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications

Quantity: 4 systems

Why Was This Recalled?

Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.

Where Was This Sold?

This product was distributed to 2 states: OH, WA