Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Digital Diagnost Stationary radiographic system Recalled by Philips Healthcare Inc. Due to When the operator for a wall stand view...

Date: October 30, 2012
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips Digital Diagnost Stationary radiographic system

Quantity: 360

Why Was This Recalled?

When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Healthcare Inc.

Philips Healthcare Inc. has 36 total recalls tracked by RecallDetector.