Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031 Recalled by Philips Healthcare Inc. Due to An inspection of IATD (Installation Acceptance Test Document)IA...

Date: January 16, 2013
Company: Philips Healthcare Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Healthcare Inc. directly.

Affected Products

Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics

Quantity: 116 units

Why Was This Recalled?

An inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment

Where Was This Sold?

This product was distributed nationwide across all 50 states.