Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips IntelliVue Module Measurement X2 Model: M3002A Recalled by Philips Electronics North America Corporation Due to The ST elevation alarm on the Patient Monitor...

Date: January 4, 2016
Company: Philips Electronics North America Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Electronics North America Corporation directly.

Affected Products

Philips IntelliVue Module Measurement X2 Model: M3002A

Quantity: 13,300

Why Was This Recalled?

The ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor

Where Was This Sold?

This product was distributed nationwide across all 50 states.