Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Mobile Detector Holder - Stationary X-ray System (RF System). suitable Recalled by Philips Electronics North America Corporation Due to If the lock mechanism of the movable camp...

Date: July 14, 2017
Company: Philips Electronics North America Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Electronics North America Corporation directly.

Affected Products

Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.

Quantity: 412

Why Was This Recalled?

If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.

Where Was This Sold?

412 units affected.