Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IntelliVue X3 Patient Monitor. Recalled by Philips Electronics North America Corporation Due to The NBP measurement of Intellivue X3 Patient Monitor...

Date: November 17, 2017
Company: Philips Electronics North America Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Electronics North America Corporation directly.

Affected Products

IntelliVue X3 Patient Monitor.

Quantity: 975 (updated 8-14-18)

Why Was This Recalled?

The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.