Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ingenia MR-OR Recalled by Philips Electronics North America Corporation Due to Possibility that a patient or user may be...

Date: December 1, 2016
Company: Philips Electronics North America Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Electronics North America Corporation directly.

Affected Products

Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite

Quantity: 9

Why Was This Recalled?

Possibility that a patient or user may be exposed to a very small level of touch current.

Where Was This Sold?

This product was distributed to 2 states: CA, NY

Affected (2 states)Not affected

About Philips Electronics North America Corporation

Philips Electronics North America Corporation has 160 total recalls tracked by RecallDetector.