Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PerkinElmer 20ul Clear RoboRack Tip Recalled by PerkinElmer, Inc Due to Defective tips have the potential to pipette/dispense volumes...

Date: October 13, 2016
Company: PerkinElmer, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PerkinElmer, Inc directly.

Affected Products

PerkinElmer 20ul Clear RoboRack Tip, Pipette (Qty 960) Item Number: 6000677. Used with the: JANUS AUTOMATED WORKSTATION Product Codes: AJS4001, AJM4001, AJI4001, AJL4001, AJS8001, AJM8001, AJI8001, AJL8001, AJM4G01, AJI4G01, AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01, AJL4M01, AJM8M01, AJI8M01, AJL8M01 NGS EXPRESS Product Codes: AJS4NGS

Quantity: 33312

Why Was This Recalled?

Defective tips have the potential to pipette/dispense volumes 50-100% lower than is specified for the tips.

Where Was This Sold?

This product was distributed to 4 states: IA, MD, NY, WA

Affected (4 states)Not affected

About PerkinElmer, Inc

PerkinElmer, Inc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report