Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Perkin Elmer RoboRack 25ul Recalled by PerkinElmer, Inc Due to Filter Tips molded incorrectly and may not seal...

Date: June 24, 2016
Company: PerkinElmer, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PerkinElmer, Inc directly.

Affected Products

Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows. The NGS Express is an automated, programmable liquid handling instrument suitable for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation workflows

Quantity: 595 trays (571,200 tips)

Why Was This Recalled?

Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing

Where Was This Sold?

This product was distributed to 7 states: CA, IA, MA, NY, OR, PA, WA

Affected (7 states)Not affected