Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 Recalled by PerkinElmer Health Sciences Canada, Inc. Due to Under certain conditions when using PerkinElmer QSight 210...

Date: September 24, 2018
Company: PerkinElmer Health Sciences Canada, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PerkinElmer Health Sciences Canada, Inc. directly.

Affected Products

PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to identify compounds in human samples by ionizing the compound and separating the resulting ions. Part Number: BC004872

Quantity: 18 units

Why Was This Recalled?

Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

Where Was This Sold?

This product was distributed to 3 states: FL, MN, PA