Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: Recalled by Percussionaire Corporation Due to Due to internal venturi component getting transiently stuck...

Date: December 23, 2024
Company: Percussionaire Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Percussionaire Corporation directly.

Affected Products

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Quantity: 4,727 packs/cases

Why Was This Recalled?

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.