Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device Recalled by Palodex Group Oy Due to It was noticed in production on 28.10.2021, that...

Date: October 28, 2021
Company: Palodex Group Oy
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Palodex Group Oy directly.

Affected Products

Orthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2D X-ray images and Xray projection images for the reconstruction of a 3D view. The device is operated and used by qualified healthcare professionals.

Quantity: 103

Why Was This Recalled?

It was noticed in production on 28.10.2021, that the final testing database gives a pass result for Pan ADC repeatability test even when the test result is not within the set acceptance criteria (repeatability within 4%).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Palodex Group Oy

Palodex Group Oy has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report