Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Recalled by Palodex Group Oy Due to Slow degradation of plastic insulation used in the...

Date: January 7, 2020
Company: Palodex Group Oy
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Palodex Group Oy directly.

Affected Products

Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Quantity: 101

Why Was This Recalled?

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

Where Was This Sold?

This product was distributed to 2 states: PA, WI