Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

e500 Automatic Transport Ventilator Recalled by O-Two Medical Technologies, Inc. Due to There is a potential for the unit to...

Date: February 12, 2019
Company: O-Two Medical Technologies, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact O-Two Medical Technologies, Inc. directly.

Affected Products

e500 Automatic Transport Ventilator, Model Number 01EVE500

Quantity: 136

Why Was This Recalled?

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Where Was This Sold?

This product was distributed to 8 states: AZ, CA, CO, FL, OH, PA, TX, VA

Affected (8 states)Not affected

About O-Two Medical Technologies, Inc.

O-Two Medical Technologies, Inc. has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report